Clear purpose: verifies switch short-circuit withstand performance, widely used in electrical manufacturing, power operation, research institutes and universities.
Power-Frequency Impulse High-Current Test System
A power-frequency high-current system for switch short-circuit withstand tests, outputting 600 A–20 kA (rms) at 240 V / 420 V with ±5% error. Compliant with IEC 62055-31 and ANSI C12.1, with power factor per IEC 61008-1.
What this solution covers
High-current output: 600 A–20 kA (rms) power-frequency short-circuit current, 240 V / 420 V test voltage, with ±5% error for both current and voltage.
Standards compliance: meets IEC 62055-31 (Uc2/Uc3/Uc4) and ANSI C12.1, with power factor per IEC 61008-1 Table 16.
Technical details & configuration
Source: domestic high-voltage test equipment (distributed by Linkall). The specifications below are taken from the manufacturer's power-frequency impulse high-current test system page; the specific current class, test voltage and supply configuration remain subject to project confirmation. See the notes in each row for the applicable scope.
| Item | Published information | Source |
|---|---|---|
| Application | Mainly for switch short-circuit withstand tests, widely used in electrical manufacturing, power operation departments, research institutes and universities | [1] |
| Applicable Standards | The high-current short-circuit test system meets IEC 62055-31 Uc2, Uc3, Uc4 and ANSI C12.1 requirements | [1] |
| Max Output Current | Max 20 kA power-frequency short-circuit current | [1] |
| Short-Circuit Test Current Range | 600 A – 20 kA (rms) | [1] |
| Test Voltage Class | 240V、420V | [1] |
| Test Error | Test current error: ±5%; test voltage error: ±5%; current ≤±2%; voltage ≤±1.5% | [1] |
| Power Factor | 600 A – 20 kA complies with IEC 61008-1 Table 16, deviation ≤-0.05 | [1] |
Provide the DUT type (switch, circuit breaker, electricity meter, etc.), short-circuit test current (600 A–20 kA rms), test voltage class (240 V or 420 V), power factor requirement, test duration and operation sequence, plus control system and calibration needs. Confirm with the manufacturer.
Questions before you start
Do the listed values apply to every configuration?
Source: domestic high-voltage test equipment (distributed by Linkall). The specifications below are taken from the manufacturer's power-frequency impulse high-current test system page; the specific current class, test voltage and supply configuration remain subject to project confirmation. See the notes in each row for the applicable scope.
What should I prepare for a technical discussion?
Provide the DUT type (switch, circuit breaker, electricity meter, etc.), short-circuit test current (600 A–20 kA rms), test voltage class (240 V or 420 V), power factor requirement, test duration and operation sequence, plus control system and calibration needs. Confirm with the manufacturer.
How do I request documentation?
Email linkall@inlinkall.com with the product name and test requirements, or call +86 21 52831768. The online enquiry form is not yet available. Documentation is supplied in its available language.