Common-mode immunity (4-16): Per IEC 61000-4-16 for AC/DC power, control and signal ports; the NSG 4060 + NSG 4060-1 + CN-series coupling networks meet the requirements (for ports with connecting cables longer than 20 m).
IEC 61000-4-16 / -4-19 测试方案
For the test plans according to IEC 61000-4-16 (0 Hz–150 kHz common-mode disturbance immunity) and IEC 61000-4-19 (2 kHz–150 kHz differential-mode disturbance immunity), the core is the NSG 4060 test system. Together with the NSG 4060-1, CN series coupling networks, CDND series coupling-decoupling networks, and CT419-5 current probe, it meets the testing requirements for common-mode voltage, differential-mode voltage, and differential-mode current respectively.
What this solution covers
Differential-mode voltage (4-19): Per IEC 61000-4-19 for AC power ports; the NSG 4060 + CDND coupling/decoupling networks meet the voltage test requirements, supporting CW pulses with pause and rectangular modulated pulses.
Differential-mode current (4-19): The NSG 4060 + CT419-5 current probe meets IEC 61000-4-19 differential-mode current testing over 2–150 kHz.
Wide signal capability: Covers DC/16 Hz to 150 kHz for short-term and sweep/continuous tests, with phase synchronisation (0° ±5%), low harmonic distortion and a range of output voltages/currents.
Technical details & configuration
The source material is a test-solution description listing frequency ranges, output voltages and impedance for each configuration. Values are not published on this page unless verified. See notes in each row.
| Item | Published information | Source |
|---|---|---|
| Core system | IEC 61000-4-16 (0 Hz–150 kHz common mode); IEC 61000-4-19 (2 kHz–150 kHz differential mode) | [1] |
| Common-mode setup (4-16) | NSG 4060 + NSG 4060-1 + CN networks; short-term DC/16 Hz–200 Hz, sweep/continuous 13 Hz–150 kHz | [1] |
| Common-mode output (4-16) | Short-term 0.1 Vrms to ≥300 Vrms (max 2 A), 1–100 V DC; sweep/continuous 0.1 Vrms to ≥30 Vrms / V DC | [1] |
| Differential voltage (4-19) | NSG 4060 + CDND networks; 2–150 kHz; 0.1 Vrms to ≥20 Vrms; modulation 3 Hz–10 kHz, 50% duty cycle | [1] |
| Differential current (4-19) | NSG 4060 + CT419-5 current probe, 2–150 kHz | [1] |
please provide the test standard and edition (4-16/4-19), port type (power/control/signal), required test levels and coupling network (CN / CDND / CT419-5) configuration; DC power-port testing per 4-16 requires the NSG 4060-op1 — configuration and availability must be confirmed per project.
Questions before you start
Do the listed values apply to every configuration?
The source material is a test-solution description listing frequency ranges, output voltages and impedance for each configuration. Values are not published on this page unless verified. See notes in each row.
What should I prepare for a technical discussion?
please provide the test standard and edition (4-16/4-19), port type (power/control/signal), required test levels and coupling network (CN / CDND / CT419-5) configuration; DC power-port testing per 4-16 requires the NSG 4060-op1 — configuration and availability must be confirmed per project.
How do I request documentation?
Email linkall@inlinkall.com with the product name and test requirements, or call +86 21 52831768. The online enquiry form is not yet available. Documentation is supplied in its available language.
Legacy information may not reflect current availability. Confirm the final model, options and specifications in the technical agreement.