Comprehensive standards: complies with IEC 61000-4-6 / GB T17626.6, CISPR 24 / GB T17618, CISPR 14-2 / GB 4343.2 and IEC 61547 / GB T18595, evaluating conducted disturbance immunity from RF fields in the 9 kHz–80 MHz range.
RF Electromagnetic Field Radiated Immunity (CS) Test Solution
The CS solution evaluates conducted disturbance immunity per IEC 61000-4-6 and related standards. In the 150 kHz–80 MHz band it applies four severity levels (1–10 V) using CDN, electromagnetic clamp, current clamp or direct injection, supported by a signal generator, attenuator and coupling/decoupling devices.
What this solution covers
Clear severity levels: four levels 1/2/3/X (1–10 V, X open) for 150 kHz–80 MHz based on installation environment; no measurement required in 9 kHz–150 kHz; small EUTs can be extended to 230 MHz.
Four injection methods: CDN, electromagnetic clamp, current clamp and direct injection, supported by a signal generator, 6 dB attenuator and coupling/decoupling devices for power lines and multi-core cables.
Technical details & configuration
Source: Linkall's system EMC test solution. The specifications below are taken from the manufacturer's RF electromagnetic field conducted immunity (CS) test solution page; the standard version, severity level selection, EUT dimensions and cable configuration remain subject to project confirmation. See the notes in each row for the applicable scope.
| Item | Published information | Source |
|---|---|---|
| Applicable Standards | IEC61000-4-6 / GB T17626.6;CISPR24 / GB T17618;CISPR14-2 / GB 4343.2;IEC61547 / GB T18595 | [1] |
| Test Objective | RF transmitter fields at 9 kHz–80 MHz act on power, communication and interface cables, which act as passive antennas coupling disturbance into the equipment; establishes a common reference for evaluating immunity to RF-field-induced conducted disturbances | [1] |
| Severity Level Rules | No measurement required for 9 kHz–150 kHz; 150 kHz–80 kHz severity selected per final installation environment; equipment <0.4 m with no conducting cable (battery-powered, ungrounded, not charging) is exempt, but must be tested if used while charging | [1] |
| Test Band & Extension | Standard 150 kHz–80 MHz; actual band determined per EUT, extendable up to 230 MHz for small EUTs | [1] |
| Severity Levels 150 kHz–80 MHz | Levels 1/2/3/X; RMS voltage U0: 120/130/140 dBμV corresponding to 1/3/10 V; X is an open level (specific) | [1] |
Provide the product category, the standard to meet, EUT dimensions and cable types, installation electromagnetic environment level, test band and whether to extend to 230 MHz, injection method selection, plus the signal generator and coupling/decoupling device configuration needs. Confirm with the manufacturer.
Questions before you start
Do the listed values apply to every configuration?
Source: Linkall's system EMC test solution. The specifications below are taken from the manufacturer's RF electromagnetic field conducted immunity (CS) test solution page; the standard version, severity level selection, EUT dimensions and cable configuration remain subject to project confirmation. See the notes in each row for the applicable scope.
What should I prepare for a technical discussion?
Provide the product category, the standard to meet, EUT dimensions and cable types, installation electromagnetic environment level, test band and whether to extend to 230 MHz, injection method selection, plus the signal generator and coupling/decoupling device configuration needs. Confirm with the manufacturer.
How do I request documentation?
Email linkall@inlinkall.com with the product name and test requirements, or call +86 21 52831768. The online enquiry form is not yet available. Documentation is supplied in its available language.
- 凌茂:射频电磁场辐射抗扰度测试系统(CS)/ LinkAll:RF Electromagnetic Field Radiation Immunity Test System (CS)Legacy website reference
Legacy information may not reflect current availability. Confirm the final model, options and specifications in the technical agreement.